Time matters as much as temperature

Classic burn research by Moritz and Henriques in the 1940s established that skin injury depends on both temperature and duration. At about 44 °C (111 °F), injury took roughly six hours of continuous contact. Above that, the time needed fell quickly with each additional degree. This is why low-temperature burns happen. A surface that feels only pleasantly warm can cause harm when a person sleeps on it or cannot feel it.

Who is most exposed

  • People with reduced sensation in the skin.
  • People who cannot move away, including sleepers and sedated patients.
  • Skin under pressure, where blood flow that would carry heat away is reduced.

Published guidance

ISO 13732-1 gives burn thresholds for contact with hot surfaces by material and contact time. Product standards in several fields set their own surface limits for parts that touch the body for long periods. We refer to the published figures relevant to each product and name them in the report.

How we apply limits

LocationApproach
Any surface that can rest on skinThe lowest limit, set for long contact
Surfaces separated from skin by layersLimit set at the skin side, verified by measurement
Element face pointing into an air gapHigher limit, with proof that the skin side stays low
Surfaces on equipment, not peopleLimits set by the materials and the application

Limits are set per measured point. One number for a whole product is either too cautious to be useful or too loose to be sound.

Why the outward heat path helps

In the HEATJAC stack, the warming element faces away from the body with insulation and a heat-reflective layer between it and the wearer. The warm surface is not the one near the skin. The element can do its work at a temperature that would be unsuitable for direct contact, while the skin side stays far lower. See heat off the skin.

Verifying

We measure the skin side directly, in the worst case: element covered, outer layer closed, highest setting, long run. A limit that holds only in the easy case is not a limit.

This page describes engineering practice. It is not medical advice.