Method: thermal imaging
A thermal camera turns the infrared radiation leaving a surface into a temperature map. It shows the whole surface at once, which no contact sensor can do. It also has limits that are easy to forget.
In this section
- ME-01 How we test
- ME-02 Method: thermal imaging
- ME-03 Method: contact temperature sensing
- ME-04 Method: power and energy measurement
- ME-05 Method: warm-up curves
- ME-06 Method: steady state and stability
- ME-07 Method: temperature uniformity
- ME-08 Method: side-by-side comparison
- ME-09 Method: cold conditions testing
- ME-10 Method: surface temperature limits
- ME-11 Method: repeatability and uncertainty
- ME-12 Standards we reference
- ME-13 Reading a thermal report
What the camera sees
Thermal cameras used in product testing sense long-wave infrared, roughly 8 to 14 micrometers. Each pixel reports the radiation arriving from one small patch of surface. Software converts that to temperature using an assumed emissivity.
What it is good for
- Distribution. Where heat is, and where it is not.
- Hot spots and cold gaps. Faults that a single sensor would miss.
- Change over time. A sequence of images shows heat spreading.
- Comparison. Two samples in one frame, under the same conditions.
What it cannot do
It sees surfaces only. It cannot see through fabric, into an air gap or under a layer. To know what is happening inside a stack, we use contact sensors.
It reads radiation and not temperature. Two surfaces at the same temperature can appear different if their emissivity differs.
Emissivity
Most fabrics, skin and painted surfaces have high emissivity, around 0.9 or above, and read close to true. Bare metals and metallized films have low emissivity. They emit little and mirror their surroundings, so a camera reads them far too low and may show a reflection of the operator. For those surfaces we apply a patch of known high-emissivity material or rely on contact sensors.
How we use it
| Practice | Reason |
|---|---|
| Fixed camera distance and angle | Images are comparable between runs |
| Emissivity set for the surface and recorded | The reading can be checked |
| Background kept uniform | Reflections are controlled |
| Regions of interest defined in advance | Averages, maximum and minimum are taken from the same areas each time |
| Images time-stamped with the sensor and power record | One timeline |
Accuracy in practice
Typical uncooled cameras are specified to about plus or minus 2 °C or 2% of reading. They are much better at showing differences across one image than at giving an absolute temperature. We use the camera for the map and contact sensors for the number.
More in the note what thermal cameras get wrong and the Knowledge article infrared thermography explained.